Composants mesurés:
Phosphine 0,01/a
Phosphine 0,1/C
Phosphine 25/a
Phosphine 50/a
Gamme de mesure:
0,01 à 1 ppm
0,1 à 3 ppm
25 à 10 000 ppm
50 à 1000 ppm
Plage d’humidité:
40 mg/L (max.)
Moins de 20 mg/L
Moins de 30 mg/L
Moins de 40 mg/L
Nb de courses:
1/10
1/3
3
3/10
4 product options available for selection
Product selection table with 4 available options. Use arrow keys to navigate and Enter or Space to select.
Numéro de catalogue. Composants mesurés Gamme de mesure Plage d’humidité Nb de courses
19822161 Phosphine 0,1/C 0,1 à 3 ppm 40 mg/L (max.) 1/3
17985290 Phosphine 0,01/a 0,01 à 1 ppm Moins de 20 mg/L 3/10
17985292 Phosphine 25/a 25 à 10 000 ppm Moins de 30 mg/L 1/10
19018612 Phosphine 50/a 50 à 1000 ppm Moins de 40 mg/L 3
Use arrow keys to navigate between rows. Press Enter or Space to select a product option. 4 options available.
4 options
Numéro de catalogue. 19822161 Fournisseur Dräger™ Code fournisseur 8103711

Pour de l'aide SVP utiliser la function de clavardage; nous envoyer un couriel à help@thermofisher.com ou appelez le service à la clientèle au 1-800-234-7437.

Short-term Dräger-Tube for the detection of phosphine, used in applications such as fertilizer fires or fumigation.

Dräger tubes are glass vials containing a chemical reagent designed to detect and measure gas concentrations. These tubes are used with manual pumps, such as the Dräger Accuro™ bellows pump, or automatic pump systems.

  • Gas Measured: Phosphine
  • Standard Measuring Range: 0.5 to 3 ppm
  • Number of Strokes (n): 1
  • Measurement Duration: Approx. 1 min
  • Standard Deviation: ±10% to 15%
  • Color Change: Yellow to red
  • Sample Volume: 100 mL
  • Temperature Range: 2°C to 40°C
  • Humidity Limit: Max. 40 mg H2O/L
  • Reaction Principle: PH3 + HgCl2 → Hg-phosphide + HCl; HCl + pH-indicator → red reaction product
  • Cross Sensitivity: Up to 6 ppm sulfur dioxide or 15 ppm hydrogen chloride do not interfere with the reading. Higher concentrations may cause positive errors. Ammonia (>100 ppm) may cause negative errors. Arsine and hydrogen sulfide are indicated with different sensitivities. 30 ppm hydrogen cyanide does not interfere.

  1. L’extension de la plage de mesure indiquée est possible pour certains tubes détecteurs. Appelez Fisher Scientific™ le 1-800-772-6733 pour plus d’informations.
  2. Certifié Safety Equipment Institute (SEI).

Spécifications

Aperçu général Tubes détecteurs de gaz à court terme
Composants mesurés Phosphine 0,1/C
Catégorie Gas Detection Tube
Plage d’humidité 40 mg/L (max.)
Gamme de mesure 0,1 à 3 ppm
Nb de courses 1/3
Quantité 10
Temps de réponse 1 à 2,5 min.
Plage de températures (impériale) 35,6 °F à 104 °F
Plage de températures (métrique) ° 2 à 40 °C