Dräger™ Tubes détecteurs à court terme - De tube Mercaptan
Description
Dräger short-term detector tubes for the measurement of Mercaptan gas concentrations.
- Standard measuring range: 0.5 to 5 ppm
- Number of strokes (n): 20
- Measurement time: approx. 5 min.
- Standard deviation: ±10 to 15 %
- Color change: White → Yellow
- Temperature range: 10°C to 40°C
- Absolute humidity: 3 to 15 mg H2O / L
- Reaction principle: 2 R-SH + Pd2+ → Pd(RS)2 + 2 H+
- Cross sensitivity: Higher molecular weight alkyl mercaptans (e.g., propyl- and butylmercaptans) are indicated with approximately the same sensitivity.
- Interfering gases: 1,000 ppm ethylene, 2,000 ppm carbon monoxide and 200 ppm hydrogen sulfide do not affect the indication.
- Pre-layer discoloration: Hydrogen sulfide discolors the pre-layer black.
- L’extension de la plage de mesure indiquée est possible pour certains tubes détecteurs. Appelez Fisher Scientific™ le 1-800-772-6733 pour plus d’informations.
- Certifié Safety Equipment Institute (SEI).
Spécifications
Spécifications
| Composants mesurés | Mercaptan 0,5/a |
| À utiliser avec (équipement) | Draeger Accuro Pump |
| Catégorie | Gas Detection Tube |
| Plage d’humidité | 3 à 15 mg/L |
| Gamme de mesure | 0,5 à 5 ppm |
| Nb de courses | 20 |
| Quantité | 10 |
| Temps de réponse | 5 min. |
| Plage de températures (impériale) | 50 °F à 104 °F |
| Plage de températures (métrique) | 10° à 40 °C |