Learn More
Thermo Scientific™ CaptureSelect™ KappaXP RoboColumn™ Pre-packed columns
Thermo Scientific™ CaptureSelect™ RoboColumn™ are small chromatography columns used for fully automated and parallel chromatographic separations on robotic liquid handling workstations.
Supplier: Thermo Scientific™ 5943212200
Description
CaptureSelect RoboColumn™ are small chromatography columns used for fully automated and parallel chromatographic separations on robotic liquid handling workstations.
CaptureSelect KappaXP matrix recognizes the kappa light-chain of human antibodies, enabling purification of all Kappa containing Ig's, Fab, and Fab2 fragments directly from complex sources and feed streams.
Features of this affinity matrix include
- Purification of Kappa monoclonal antibodies, Fab, and Fab2 fragments in a single step with high purity and yield
- Mild elution for Fab fragments and antibodies
- Human-specific, no binding to bovine antibodies
- Excellent scalability
- Excellent pH stability up to 100 mM NaOH (limited exposure)
- Non-animal derived
Free of animal components
CaptureSelect products contain affinity ligands based on recombinant single-domain antibody fragments (VHH) created by a proprietary technology. The VHH affinity ligand is a 12 to 15 kDa fragment comprising the three complementarity-determining regions (CDRs) that form the antigen-binding domain, efficiently produced in the yeast Saccharomyces cerevisiae by a production process free of any animal components (Animal Origin-Free).
Characteristics
- Column size: 200 μL, 1 strip, 8 columns (8-row)
- Column dimension: 5-mm diameter, 10-mm bed height
- Matrix: agarose based, epoxide-activated
- Average particle size: 65 ± 10 μm
- Ligand: CaptureSelect KappaXP affinity ligand
- Ligand coupling method: epoxide coupling
- Fab binding capacity: 20–30 g IgG Kappa-Fab/liter resin depending on flow rate, column height, and residence time
- IgG binding capacity: 35–45 g IgG/liter resin depending on flow rate, column height, and residence time
- Elution conditions: 20 mM citric or acetic acid, pH 3–4; 100 mM Tris, 1.5 M MgCl2, pH6
- Formulation buffer: 20% (v/v) ethanol
- Max operating pressure: 2 bar
- Flow rate: 16–150 cm/hr
RoboColumn is a trademark of Repligen.
Specifications
| Store in refrigerator (2–8°C). |
|
| 65 ±10 μm | |
| 200 μL | |
| KappaXP Affinity Matrix with KappaXP Ligand |
| Wet Ice | |
| CaptureSelect | |
| Column | |
| 8-well Strips |