Applied Biosystems™ MicroAmp™ 96-Well Tray/Retainer Set

Catalog No. 403081
Encompass
Change view
Click to view available options
Quantity:
10 Set
1 product options available for selection
Product selection table with 1 available options. Use arrow keys to navigate and Enter or Space to select.
Catalog No. Quantity
403081 10 Set
Use arrow keys to navigate between rows. Press Enter or Space to select a product option. 1 options available.
1 options
Catalog No. 403081 Supplier Applied Biosystems™ Supplier No. 403081
Only null left

Includes

Holds up to 96 MicroAmp tubes w/o caps (0.2mL) for processing either on the System 9600 or 9700. Use with MicroAmp 96-Well Base P/N N8010531

Available in 96- or 24-well format

This set holds MicroAmp reaction tubes without caps for processing on the Veriti 96-Well Thermal Cycler, 96-Well GeneAmp PCR System 9700, or the Applied Biosystems 2720 Thermal Cycler.

  • Available in 96- or 24-well format, offering flexibility
  • Designed for compatibility with MicroAmp reaction tubes and MicroAmp8-Strip reaction tubes
  • Organize tubes while adding samples and mixing
  • Reduces spillage and tube mix-ups by securing and ordering tubes on lab bench
  • When samples are ready, tray of tubes are put directly on block of GeneAmp Thermal Cycler
  • Tray eliminates hassle of transferring tubes and provides thermal contact necessary for high-performance thermal cycling

See user's manual or package insert for limited label license, and trademark information.

PCR, PCR and Real-Time PCR

Order Info

Shipping Condition: Room temperature

Specifications

Certifications/Compliance Certified DNA/RNase Free
Content And Storage Holds up to 96 MicroAmp™ tubes w⁄o caps (0.2 mL) for processing either on the System 9600 or 9700. Use with MicroAmp™ 96-Well Base P⁄N N8010531
Product Line MicroAmp
Color White
Product Type 96-Well Tray/Retainer Set
For Use With (Equipment) 2720 Thermal Cycler, 7000 System, AB 2720 Thermal Cycler, GeneAmp 9700, 2720 Thermal Cycler
For Use With (Application) PCR
Plate Compatibility 96-Well Plates
Quantity 10 Set
Shipping Condition Room Temperature
Why are my PCR tubes crushed/deformed after my PCR run?

The heated thermal cycler lid is designed to make sure that optimized pressure is applied on the microplate for an efficient PCR reaction. When using PCR tubes, excess pressure on the tubes, from the heated lid can cause deformation of the tubes. To avoid this, we recommend using the Tray/Retainer set, as shown below:

Cat. No. 4381850, MicroAmp 96-Well Tray/Retainer Set for Veriti Systems – This Tray is for use with MicroAmp Strip Tubes (or MicroAmp Reaction Tubes without Caps, 0.2 mL), with the Veriti Thermal Cycler, SimpliAmp Thermal Cycler, and Proflex PCR System.
Cat. No. 403081, MicroAmp 96-Well Tray/Retainer Set – This Tray is for use with MicroAmp Strip Tubes (or MicroAmp Reaction Tubes without Caps, 0.2 mL), with the 7000 System, 2720 Thermal Cycler, and GeneAmp PCR System 9700.
Cat. No. 4379983, MicroAmp 96-Well Tray for VeriFlex Blocks – This Tray is for use with MicroAmp Strip Tubes (or MicroAmp Reaction Tubes with Caps, 0.1 mL), with the Veriti Thermal Cycler, Veriti Fast Thermal Cycler, SimpliAmp Thermal Cycler, and Proflex PCR System.

Please take a look at this video: How to Use Adapter Tray Retainers in Thermal Cycler Starter Kits (https://www.youtube.com/watch?v=1uh2ZxZJtgs).

What does the SeqStudio Starter Kit (PN A35000) contain?

The SeqStudio Starter Kit (PN A35000) includes:
- SeqStudio Integrated Capillary Protector (Cat. No. A31923)
- SeqStudio Cathode Buffer Container (Cat. No. A33401)
- Sequencing Standard, BigDye Terminator v3.1, for 3500/SeqStudio Flex (Cat. No. 4404312)
- DS-33 GeneScan Installation Standards with GeneScan 600 LIZ Size Standard v2.0 (Cat. No. 4376911)
- Hi-Di Formamide (Cat. No. 4401457)
- MicroAmp Optical 96-Well Reaction Plate with Barcode (Cat. No. 4306737)
- MicroAmp 96-Well Tray/Retainer Set (Cat. No. 403081)
- MicroAmp Optical 8-Tube Strip, 0.2 mL (Cat. No. 4316567)
- Nuclease-Free Water (not DEPC-Treated) (Cat. No. AM9930)
- Cathode Buffer Container Reservoir Septa for SeqStudio Genetic Analyzer (Cat. No. A35640)
- 96-well Septa for SeqStudio Genetic Analyzer (Cat. No. A35641)
- 8-strip Septa for SeqStudio Genetic Analyzer (Cat. No. A35643)
- SeqStudio Cartridge v2 (PN A41328) (if this component is ordered online the catalog number is A41331)