Thermo Scientific™ GENESYS™ 50 UV-Vis Spectrophotometer PROMO

For teaching labs or quality control experts who require ruggedness, ease-of-use and software capability in UV-Vis spectrophotometry.

Supplier:  Thermo Scientific™ 8402980001PM24

Catalog No. 143853511P4

Please call Customer Service at 1-800-234-7437 or send an email to help@thermofisher.com for assistance.


Description

Description

The GENESYS 50 is a dual beam UV-Vis spectrophotometer in a single-cell configuration that meets all of the specifications for critical quality control applications and precise chemical measurements. The GENESYS 40 Dual Beam Vis Spectrophotometer is a visible spectral region version in this instrumental class. Both feature data resolution from 0.2 nm to 5 nm.

Features:
  • 7-inch high-resolution touchscreen user interface
  • Local control for speed and reliability
  • Networking capable (Wi-Fi adaptor available)
  • Large sample compartment with front access
Designed for:
  • Chemistry labs
  • Routine measurements
  • Moderate throughput requirements
Specifications

Specifications

±0.002 A
Single USB-A; Duplex USB-A; Export data to network or PC via USB, Ethernet, or Wi-Fi USB adaptor, available; Print via USB, Ethernet or Wi-Fi USB adaptor available
Silicon Photodiode
35.5 x 38.5 x 19.5 cm
190 to 1100 nm
External AC to DC converter. Voltage and Frequency (Hz) selected automatically, 100/240 V, 50/60 Hz
>5 years typical, 3 years guaranteed
Dual Beam
-3 A to +5 A
±0.001 A at 1 A measured at 1.0 A at 546 nm
Slow, medium and fast - up to 1600 nm/min.
2 years standard with 1 year extension upon registration of the instrument with Thermo Fisher Scientific within 6 months of purchase
0.2 nm, 0.5 nm, 1 nm, 2 nm, 5 nm
7.5 kg
Accessible from top, front or side; Able to accommodate cells up to 100 mm cells; Available accessory for test-tubes up to 25 mm diameter and 150 mm height; Removable, washable sample compartment liner with magnetic placement and hold-down
GENESYS 50 UV-Vis Spectrophotometer
Touchscreen
2 nm
(At 500 nm after 1 hour warmup) <0.0005 A/hr.
Touchscreen
(RMS at 500 nm 60 consecutive measurements): ≤0.00020A at 0A at 260 and 500 nm, ≤0.00030A at 1A at 260 and 500 nm, ≤0.00040A at 2A at 260 and 500 nm
±0.002A at 0.5A, ±0.004A at 1.0A, ±0.008 A at 2.0 A
-2 A to +3.5 A
Snap-on printer available
< 1.0%T 198 nm (KCl) , <0.05%T at 220 nm (NaI), <0.03%T at 340 nm (NaNO2)
±0.5 nm
< ±0.2 nm
SDS
Documents

Documents

Promotions